Remote Control Commands
R&S
®
SMU200A
676Operating Manual 1007.9845.32 ─ 15
Parameters:
float
-80 dB to 0 dB
0.01 dB
0 dB
<Start>
Range:
Increment:
*RST:
BB:MCCW:EDIT:CARR:POW -50 dB
sets the power of the carrier to -50 dB.
Example:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STARt <CarrierIndex>
The command selects the first carrier in the carrier range to which the settings with
the :BB:MCCW:EDIT:CARR:... commands shall apply.
Parameters:
float
0
<Start>
*RST:
BB:MCCW:EDIT:CARR:STAR 4
the carrier range starts at carrier 4.
Example:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STATe <State>
The command switches all the carriers in the selected carrier range on or off.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
BB:MCCW:EDIT:CARR:STAT ON
sets all the carriers in the carrier range to ON.
Example:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:STOP <Stop>
The command selects the last carrier in the carrier range to which the settings with
the :BB:MCCW:EDIT:CARR:... commands shall apply.
Parameters:
float
0
<Stop>
*RST:
BB:MCCW:EDIT:CARR:STOP 40
the carrier range stops at carrier 40.
Example:
Trigger Settings
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute.........................................................677
[:SOURce<hw>]:BB:MCCW:TRIGger:EXECute.................................................................677
SOURce Subsystem