Remote Control Commands
R&S
®
SMU200A
677Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:DELay.................................................677
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:INHibit.................................................678
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe..................................................................678
[:SOURce<hw>]:BB:MCCW:TRIGger:SLENgth.................................................................679
[:SOURce<hw>]:BB:MCCW:TRIGger:SOURce.................................................................679
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:DELay.............................................679
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:INHibit.............................................680
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:SYNChronize:OUTPut......................680
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence.............................................................680
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute
Stops signal generation for trigger modes "Armed Auto" and "Armed Retrigger". A sub-
sequent internal or external trigger event restart signal generation.
BB:MCCW:TRIG:ARM:EXEC
stops signal generation for trigger modes "Armed Auto" and
"Armed Retrigger".
Example:
EventUsage:
[:SOURce<hw>]:BB:MCCW:TRIGger:EXECute
Executes a trigger. The internal trigger source must be selected using the command
SOUR:BB:DAB:TRIG:SOUR INT and a trigger mode other than "AUTO" must be
selected using the command SOUR:BB:MCCW:TRIG:SEQ.
BB:MCCW:TRIG:SOUR INT
sets internal triggering.
BB:MCCW:TRIG:SEQ RETR
sets Retrigger mode, i.e. every trigger event causes signal gen-
eration to restart.
BB:MCCW:TRIG:EXEC
executes a trigger.
Example:
EventUsage:
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:DELay <Delay>
The command specifies the trigger delay for triggering by the trigger signal from the sec-
ond path (two-path instruments only).
Parameters:
0 to 2^32-1
0
<Delay>
Range:
*RST:
SOURce Subsystem