Remote Control Commands
R&S
®
SMU200A
678Operating Manual 1007.9845.32 ─ 15
BB:MCCW:TRIG:SOUR OBAS
sets for path A the internal trigger executed by the trigger signal
from the second path (path B).
BB:MCCW:TRIG:OBAS:DEL 50
sets a delay of 50 symbols for the trigger.
Example:
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:INHibit <Inhibit>
The command specifies the number of samples/symbols by which a restart is to be
inhibited following a trigger event. This command applies only for triggering by the second
path (two-path instruments only).
Parameters:
0 to 2^26-1
0
<Inhibit>
Range:
*RST:
BB:MCCW:TRIG:SOUR OBAS
sets for path A the internal trigger executed by the trigger signal
from the second path (path B).
BB:MCCW:TRIG:INH 200
sets a restart inhibit for 200 symbols following a trigger event.
Example:
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe
Queries the current status of signal generation for all trigger modes with modulation on.
Parameters:
STOP|RUN
STOP
The signal is generated. A trigger event occurred in the triggered
mode.
RUN
The signal is not generated. A trigger event did not occur in the
triggered modes, or signal generation was stopped by the
command BB:MCCW:TRIG:ARM:EXECute (armed trigger modes
only).
RST value
<Rmode>
*RST:
BB:MCCW:TRIG:SOUR EXT
sets external triggering via the TRIGGER connector.
BB:MCCW:TRIG:MODE ARET
selects the Armed_Retrigger mode
BB:DAB:TRIG:RMOD?
queries the current status of signal generation.
Response: Run
the signal is generated, an external trigger was executed.
Example:
SOURce Subsystem