Remote Control Commands
R&S
®
SMU200A
684Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:ONTime <OnTime>
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:OFFTime <OffTime>
Sets the number of symbols/samples in a period (ON time + OFF time) during which the
marker signal in setting SOURce:BB:MCCW:TRIGger:OUTPut:MODE RATio on the
marker outputs is OFF.
Parameters:
1 to 2^24-1
1
1
<OffTime>
Range:
Increment:
*RST:
BB:MCCW:TRIG:OUTP2:OFFT 2000
sets an OFF time of 2000 symbols/samples for marker signal 2.
Example:
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:PATTern <Pattern>
Defines the bit pattern used to generate the marker signal.
Parameters:
<Hex/Binary pattern><Pattern>
BB:MCCW:TRIG:OUTP2:PATT #H39FE0000,32
sets the bit pattern.
BB:MCCW:TRIG:OUTP2:MODE PATT
activates the marker signal according to a bit pattern on output
MARKER 2.
Example:
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:PULSe:DIVider <PlsDivider>
The command sets the divider for Pulse marker mode
(SOUR:BB:MCCW:TRIGr:OUTP:MODE PULSe). The resulting pulse frequency is
derived by dividing the symbol/sample rate by the divider.
Parameters:
2 to 1024
1
2
<PlsDivider>
Range:
Increment:
*RST:
BB:MCCW:TRIG:OUTP2:PULS:DIV 2
sets the divider to 2 for the marker signal on output MARKER 2.
BB:MCCW:TRIG:OUTP2:FREQ?
queries the resulting pulse frequency of the marker signal.
Response: 66 000
the resulting pulse frequency is 66 kHz.
Example:
SOURce Subsystem