Remote Control Commands
R&S
®
SMU200A
713Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:FREQuency[:CW|:FIXed] <Fixed>
The command sets the frequency of the RF output signal for CW mode
(SOURce:FREQuency:MODE CW). In Sweep mode (SOURce:FREQuency:MODE
SWEep), this value is linked to the current sweep frequency.
In addition to a numerical value, it is also possible to specify UP and DOWN. The frequency
is then increased or decreased by the value which is set under
[SOURce<[1]|2>:]FREQuency:STEP.
As with the FREQ value entered in the display, the OFFSet value is also taken into con-
sideration with this command. The specified value range is therefore only effective if
OFFSet is set to 0. The value range for other OFFset values can be calculated using
the following formula:
300 kHz + OFFSet ... RF
max
+ OFFSet
Correlation: FREQ for FREQ:MODE SWE linked to sweep frequency.
Parameters:
float
300kHz to RFmax
0.01Hz
100 MHz
Hz
<Fixed>
Range:
Increment:
*RST:
Default unit:
FREQ 500kHz
sets the frequency of RF output signal A to 500 kHz.
Example:
SCPI conformUsage:
[:SOURce<hw>]:FREQuency[:CW|:FIXed]:RCL <Rcl>
The command determines whether the current frequency setting is retained or whether
the stored frequency setting is adopted when an instrument configuration is loaded.
*RST does not affect this setting.
In case of two-path instruments, the setting is valid for both paths. If a suffix is specified,
it is ignored.
Parameters:
INCLude|EXCLude
INCLude
The stored frequency is also loaded when a stored instrument
configuration is loaded.
EXCLude
The RF frequency is not loaded when a stored instrument
configuration is loaded. The current frequency is retained.
no default
<Rcl>
*RST:
FREQ:RCL INCL
The stored frequency is set if the Recall command is called.
Example:
SOURce Subsystem