Remote Control Commands
R&S
®
SMU200A
714Operating Manual 1007.9845.32 ─ 15
SCPI conformUsage:
[:SOURce<hw>]:FREQuency:MANual <Manual>
In Sweep mode (:SOUR:FREQ:MODE SWE) the command sets the frequency for the next
sweep step in the Step sweep mode (SOUR:SWE:MODE MAN). Here only frequency val-
ues between the settings [SOUR]:FREQ:STAR and ...:STOP are permitted. Each
sweep step is triggered by a separate SOUR:FREQ:MAN command.
As with the Frequency value entered in the header, the OFFSet value is also taken into
consideration with this command. The specified value range is therefore only effective if
OFFSet is set to 0. The value range for other OFFset values can be calculated using
the following formula.
STARt + OFFSet ... STOP +OFFSet
Parameters:
float
STARt to STOP
0.01Hz
100MHz
Hz
<Manual>
Range:
Increment:
*RST:
Default unit:
SWE:MODE MAN
sets the Step sweep mode.
FREQ:MAN 500MHz
sets an RF frequency of 500 MHz for the next step in the Step
sweep mode.
FREQ:MODE SWE
sets the Frequency Sweep mode. An RF frequency of 500 MHz is
output.
FREQ:MAN 550MHz
triggers the next sweep step with an RF frequency of 550 MHz.
Example:
Example:
SCPI conformUsage:
[:SOURce<hw>]:FREQuency:MODE <Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
SOURce Subsystem