EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 745

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMU200A
725Operating Manual 1007.9845.32 ─ 15
Parameters:
float
-10 to 10
0.02PCT
0
PCT
<I>
Range:
Increment:
*RST:
Default unit:
IQ:IMP:LEAK:I 3 PCT
sets the leakage for the I-channel to 3 percent.
Example:
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q <Q>
This command sets the carrier leakage amplitude for the Q-channel of the I/Q modulation.
Parameters:
float
-10 to 10
0.02PCT
0PCT
PCT
<Q>
Range:
Increment:
*RST:
Default unit:
IQ:IMP:LEAK:Q 3 PCT
sets the leakage for the Q-channel to 3 percent.
Example:
[:SOURce<hw>]:IQ:IMPairment:QUADrature:[ANGLe] <Angle>
This command sets the quadrature offset for the I/Q modulation.
Parameters:
float
-10.0DEG to 10.0DEG
0.02DEG
0DEG
DEG
<Angle>
Range:
Increment:
*RST:
Default unit:
IQ:IMP:QUAD:ANGL -5DEG
sets the quadrature offset to -5 degrees.
Example:
[:SOURce<hw>]:IQ:IMPairment[:STATe] <State>
The command activates (ON) and deactivates (OFF) the three impairment or correction
values LEAKage, QUADrature and IQRatio for the analog signal in the I/Q modulator.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
SOURce Subsystem