Remote Control Commands
R&S
®
SMU200A
731Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP:RESult?
Displays the resulting output off-load voltage of both signal components at the rear panel.
For two-path instruments, the outputs are assigned with command
SOURce:BB:IQOut:SOURce.
Return values:
float
0.001 V
0 V
<Result>
Increment:
*RST:
Query onlyUsage:
[:SOURce<hw>]:IQ:OUTPut:MODE <Mode>
This command selects the mode for setting the outputs.
Selection FIXed locks all IQ:OUTPut-commands except for IQ:OUTPut:TYPE and sets
them to predefined values ( *RST).
For two-path instruments, the outputs are assigned with command
[:SOURce]:BB:IQOutput:SOURce.
Parameters:
FIXed|VARiable
FIXed
The settings are fixed and cannot be changed.
Single ended: IQ:OUTP:POW:PEP = 1V.
Differential: IQ:OUTP:POW:PEP = 2V.
IQ:OUTPut:BIAS:I|Q = 0V
IQ:OUTPut:OFFSet:I|Q = 0V
VARiable
The settings can be changed (command
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:PEP
on page 730).
FIXed
<Mode>
*RST:
SOUR:IQ:OUTP:MODE VAR
SOUR:IQ:OUTP:ANAL:POW:PEP 2V
Example:
SOURce Subsystem