Remote Control Commands
R&S
®
SMU200A
732Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:IQ:OUTPut:OFFSet:I <I>
The command is only valid for output type Differential (IQ:OUTP:TYPE DIFF).
This command sets an offset between the inverting and non-inverting output.
The maximum overall voltage
(0.5 * IQ:OUTP:POW:PEP + SOUR:IQ:OUTP:BIAS:I|Q + 0.5 * SOUR:IQ:
OUTP:OFFS:I|Q) is 3.7 volts.
For two-path instruments, the outputs are assigned with command
[:SOURce]:BB:IQOutput:SOURce on page 728.
Parameters:
float
-0.3 to 0.3
0.2mV
0V
V
<I>
Range:
Increment:
*RST:
Default unit:
IQ:OUTP:TYPE DIFF
the I/Q-signal components are output differential
:IQ:OUTP:OFFS:I 0.05 V
sets an offset of 50 mV for the I-signal.
Example:
[:SOURce<hw>]:IQ:OUTPut:OFFSet:Q <Q>
The command is only valid for output type Differential (IQ:OUTP:TYPE DIFF).
This command sets an offset between the inverting and non-inverting output.
The maximum overall voltage
(0.5 * IQ:OUTP:POW:PEP + SOUR:IQ:OUTP:BIAS:I|Q + 0.5 * SOUR:IQ:
OUTP:OFFS:I|Q) is 3.7 volts.
For two-path instruments, the outputs are assigned with command
[:SOURce]:BB:IQOutput:SOURce on page 728.
Parameters:
float
-0.3 to 0.3
0.2 mV
0 V
<Q>
Range:
Increment:
*RST:
IQ:OUTP:TYPE DIFF
The I/Q-signal components are output differential.
IQ:OUTP:OFFS:I 0.05 V
sets an offset of 50 mV for the I-signal.
Example:
SOURce Subsystem

Table of Contents