Remote Control Commands
R&S
®
SMU200A
735Operating Manual 1007.9845.32 ─ 15
Parameters:
OFF|UNCHanged
OFF
<Pon>
*RST:
IQ:OUTP:DIG:PON OFF
deactivates the output when the instrument is switched on.
IQ:OUTP:DIG:PON UNCH
sets the output to the status that was active before the last switch
off.
Example:
B13 and B18Options:
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:LEVel <Level>
This command sets the rms level of the digital output signal. The level entry fields and
the appropriate units (dBFS) are displayed in the status bar. The level display refers to
both signal components (√(I
2
+Q
2
).
Parameters:
float
-90 to 0
0.0001 dB
0 dB
dBFs
<Level>
Range:
Increment:
*RST:
Default unit:
IQ:OUTP:DIG:POW:LEV -10
sets the output signal level to -10 dBFS.
Example:
B13 and B18Options:
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:PEP <Pep>
This command sets the peak envelope power of the digital output signal (PEP). The level
entry fields and the appropriate units (dBFS) are displayed in the status bar. The level
display refers to both signal components (√(I
2
+Q
2
).
Parameters:
float
-90 dBFS to 0 dBFS
0.01 dB
0 dB
dBFS
<Pep>
Range:
Increment:
*RST:
Default unit:
IQ:OUTP:DIG:POW:PEP -10
sets the signal output level to -10 dBFS.
Example:
B13 and B18Options: