Remote Control Commands
R&S
®
SMU200A
743Operating Manual 1007.9845.32 ─ 15
LFO:SWE:MODE STEP
sets the step-by-step processing of the LF sweep.
LFO:SWE:EXEC
starts a step of the LF sweep.
Example:
EventUsage:
[:SOURce]:LFOutput:SWEep[:FREQuency]:MODE <Mode>
The command sets the cycle mode of the LF sweep.
The assignment of the IEC/IEEE-bus commands to the sweep modes is given in the
description of the sweep menus.
Parameters:
AUTO|MANual|STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each frequency step of the sweep
is triggered individually, either by varying the Current Frequency
value using the rotary knob under manual control or by means of
a :LFOutput:FREQ:MAN command under remote control. With
manual control, the frequency increases or decreases (depending
on the direction of the rotary encoder) by the value specified
under :LFOutput:FREQ:STEP:INCRement. With remote
control, the frequency is set directly with the
command :LFOutput:FREQ:MAN.
STEP
Each trigger triggers one sweep step only. The frequency
increases by the value entered under
[SOURce:]LFOutput:SWEep:STEP.
AUTO
<Mode>
*RST:
STEP
Each trigger triggers one sweep step only. The frequency increa-
ses by the value entered under
[SOURce:]LFOutput:SWEep:STEP.