Remote Control Commands
R&S
®
SMU200A
754Operating Manual 1007.9845.32 ─ 15
SCPI conformUsage:
[:SOURce<hw>]:LIST:FREQuency <Frequency>
The command fills the FREQuency part of the selected list with data. The data can be
given either as a list of numbers (list can be of any length and list entries must be sepa-
rated by commas) or as binary block data. When block data is transferred, 8 (4) bytes
are always interpreted as a floating-point number with double accuracy (see the com-
mand FORMat:DATA).
*RST does not affect data lists.
Parameters:
list of numbers | block data
300 kHz to RFmax (RFmax depending on model)
<Frequency>
Range:
LIST:SEL 'D:/Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
SOUR:LIST:FREQ 1.4GHz, 1.3GHz, 1.2GHz,...
specifies the frequency values in list3. If the list already contains
data, it is overwritten.
Example:
SCPI conformUsage:
[:SOURce<hw>]:LIST:FREQuency:POINts?
The command queries the length (in points) of the FREQuency component of the selected
list.
Return values:
float<Points>
LIST:SEL 'D:/Lists/Listmode/list3'
selects list3 for editing. List3 is created if it does not yet exist.
LIST:FREQ:POIN?
queries the number of frequency values in list3.
Response: 327
list3 has 327 frequency entries.
Example:
Query onlyUsage:
[:SOURce<hw>]:LIST:INDex <Index>
The command sets the list index in step mode (LIST:MODE STEP).
Parameters:
float
0
<Index>
*RST:
SOURce Subsystem