Remote Control Commands
R&S
®
SMU200A
755Operating Manual 1007.9845.32 ─ 15
LIST:SEL 'D:/Lists/Listmode/list3'
selects list3 for use in List mode.
FREQ:MODE LIST
activates List mode. List 3 is processed.
LIST:MODE STEP
selects manual, step-by-step processing of the list.
LIST:IND 5
the frequency/level value pair with index 5 is executed.
TRIG:LIST:SOUR SING
selects triggering by means of the single trigger. The list is exe-
cuted once.
SOUR:LIST:TRIG:EXEC
triggers the processing of the selected list.
Example:
[:SOURce<hw>]:LIST:INDex:STARt <Start>
The command sets the start index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list. Only the values in the set index range
(:LIST:INDex:STARt ... :LIST:INDex:STOP) are processed in List mode.
Parameters:
float
0 to list length
0
<Start>
Range:
*RST:
LIST:SEL 'D:/Lists/Listmode/list3'
selects list3 for use in List mode.
LIST:IND:STAR 25
sets 25 as start index of the index range.
LIST:IND:STOP 49
sets 49 as stop index of the index range.
FREQ:MODE LIST
activates List mode. The frequency/level value pairs from index 25
to index 49 in list3 are processed. All other entries of the list are
ignored.
Example:
[:SOURce<hw>]:LIST:INDex:STOP <Stop>
The command sets the stop index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list. Only the values in the set index range
(:LIST:INDex:STARt ... :LIST:INDex:STOP) are processed in List mode.
Parameters:
float
0 to list length
0
<Stop>
Range:
*RST:
SOURce Subsystem