Remote Control Commands
R&S
®
SMU200A
767Operating Manual 1007.9845.32 ─ 15
Parameters:
INCLude | EXCLude
INCLude
The stored level is also loaded when a stored instrument
configuration is loaded.
EXCLude
The RF level is not loaded when a stored instrument configuration
is loaded. The current level is retained.
no default
<Rcl>
*RST:
POW:RCL INCL
the stored level is set if the Recall command is called.
Example:
SCPI conformUsage:
[:SOURce<hw>]:POWer:LIMit[:AMPLitude] <Amplitude>
The command limits the maximum RF output level in CW and SWEEP mode. It does not
influence the LEVEL display or the response to the POW? query command.
The value is not affected by an instrument preset (PRESET key), *RST and the Save/
Recall function. This parameter is influenced only by the factory preset (SYST:FPR) and
its factory value is equal to the upper limit.
Parameters:
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The value range depends on whether or not the instrument is
equipped with the option High Power.
The values are given in the data sheet.
0.01 dB
30 dBm
dBm
<Amplitude>
Increment:
*RST:
Default unit:
SOURce:POWer:LIMit:AMPLitude 10 or :POW:LIM 10
limits the RF level to maximum +10 dBm.
Example:
SCPI conformUsage:
SOURce Subsystem