Remote Control Commands
R&S
®
SMU200A
774Operating Manual 1007.9845.32 ─ 15
[:SOURce]:ROSCillator:EXTernal:RFOFf:STATe <State>
The command determines if the RF output is switched off in case of a missing external
reference signal for selection external source.
If enabled, this setting ensures that no improper RF signal due to the missing external
reference signal is output and used for measurements.
In addition to the error message "Ext Ref missing", message "RF output deactivated' is
generated.
Parameters:
0|1|OFF|ON<State>
ROSC:SOUR EXT
selects the external source. The reference must be input at the
REF IN input.
ROSC:EXT:RFOF:STAT ON
In case of a missing external signal, no RF signal is output.
Example:
Example:
[:SOURce]:ROSCillator:EXTernal:SBANdwidth <Sbandwidth>
The command selects the synchronization bandwidth for an external reference signal.
Parameters:
WIDE|NARRow
NARRow
Synchronization bandwidth is approx. 1 Hz.
WIDE
Synchronization bandwidth is approx. 750 Hz.
<Sbandwidth>
ROSC:SOUR EXT
selects the external source.
ROSC:EXT:FREQ 5 MHz
informs the instrument that the external reference has a frequency
of 5 MHz.
ROSC:EXT:SBAN WID
selects wideband setting for synchronization bandwidth.
Example:
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue <Value>
The command specifies the frequency correction value (adjustment value).
Parameters:
float
0 to 4095
1
<Value>
Range:
Increment:
ROSC:ADJ:VAL 1400
sets the adjustment value to 1400.
Example:
SOURce Subsystem