776Operating Manual 1007.9845.32 ─ 15
This example shows how to set up a frequiency sweep.
1. Set the sweep range.
[SOURce:]FREQuency:CENTer 200 MHz
[SOURce:]FREQuency:SPAN 300 MHz
2. Select linear or logarithmic spacing.
[SOURce:]SWEep[:FREQuency]:SPACing LIN
3. Set the step width and dwell time.
[SOURce:]SWEep[:FREQuency]:STEP:LINear 20 MHz
[SOURce:]SWEep[:FREQuency]:DWELl 12 ms
4. Select the trigger mode.
TRIGger:]FSWeep:SOURce SINGle
5. Select the sweep mode and activate the sweep.
[SOURce:]SWEep[:FREQuency]:MODE AUTO
[SOURce:]FREQuency:MODE SWEep
6. Trigger the sweep.
[SOURce:]SWEep[:FREQuency]:EXECute
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl..................................................................776
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute..............................................................777
[:SOURce<hw>]:SWEep[:FREQuency]:MODE..................................................................777
[:SOURce<hw>]:SWEep[:FREQuency]:POINts.................................................................778
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe.................................................................778
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing...............................................................779
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear].......................................................779
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic................................................780
[:SOURce<hw>]:SWEep:POWer:AMODe.........................................................................780
[:SOURce<hw>]:SWEep:POWer:DWELl..........................................................................781
[:SOURce<hw>]:SWEep:POWer:EXECute.......................................................................781
[:SOURce<hw>]:SWEep:POWer:MODE...........................................................................781
[:SOURce<hw>]:SWEep:POWer:POINts..........................................................................782
[:SOURce<hw>]:SWEep:POWer:SHAPe..........................................................................782
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE.............................................................783
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic].......................................................783
[:SOURce<hw>]:SWEep:RESet[:ALL]..............................................................................784
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
The command sets the time taken for each frequency step of the sweep.
The keyword [:FREQuency] can be omitted (see example). The command is then SCPI-
compliant.