Remote Control Commands
R&S
®
SMU200A
777Operating Manual 1007.9845.32 ─ 15
Parameters:
float
3ms to 10s
0.1m
10m
s
<Dwell>
Range:
Increment:
*RST:
Default unit:
SWE:DWEL 12 ms
sets a dwell time of 12 ms for a frequency sweep at the RF output.
Example:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
The command triggers a sweep.
SWE:FREQ:EXEC
triggers a frequency sweep at the RF output.
Example:
EventUsage:
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
The command sets the sweep mode.
The keyword [:FREQuency] can be omitted (see example). The command is then SCPI-
compliant.
Parameters:
AUTO|MANual|STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each frequency step of the sweep
is triggered individually, either by varying the Current Frequency
value using the rotary knob under manual control or by means of
a FREQ:MAN command under remote control. With manual
control, the frequency increases or decreases (depending on the
direction of the rotary encoder) by the value specified under
FREQ:STEP:INCRement. With remote control, the frequency is
set directly with the command :FREQ:MAN.
STEP
Each tcrigger triggers one sweep step only (Mode Single Step).
The frequency increases by the value entered under
SOUR:SWE:FREQ:STEP:LIN (linear spacing)
or ...:STEP:LOG (logarithmic spacing).
AUTO
<Mode>
*RST:
SWE:MODE AUTO
selects Mode Auto for a frequency sweep at the RF output.
Example: