Remote Control Commands
R&S
®
SMU200A
779Operating Manual 1007.9845.32 ─ 15
SOUR:SWE:SHAP TRI
selects the sweep cycle with alternating ascending and descend-
ing sweep directions.
Example:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing <Spacing>
The command selects linear or logarithmic sweep spacing.
The keyword [:FREQuency] can be omitted (see example). The command is then SCPI-
compliant.
Parameters:
LINear|LOGarithmic
LIN
<Spacing>
*RST:
SWE:SPAC LIN
selects linear sweep spacing for a frequency sweep at the RF out-
put.
Example:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] <Linear>
The command sets the step width for linear sweeps.
This command is linked to the command :SWEep[:FREQuency]:POINts as follows:
POINts = (SPAN / STEP:LIN) + 1
If STEP:LIN changes, the value of POINTs is adjusted. The STARt and STOP value is
retained.
The keywords [:FREQuency] and [:LINear] can be omitted (see example). The com-
mand is then SCPI-compliant.
Correlation: The value of :SWE:POINts is adjusted automatically.
Parameters:
float
0 to (STOP - STARt)
0.1Hz
1MHz
Hz
<Linear>
Range:
Increment:
*RST:
Default unit:
SOURce Subsystem