Remote Control Commands
R&S
®
SMU200A
782Operating Manual 1007.9845.32 ─ 15
Parameters:
AUTO|MANual|STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each level step of the sweep is
triggered individually, either by varying the Current Level value
using the rotary knob under manual control or by means of a
POW:MAN command under remote control. With manual control,
the level increases or decreases (depending on the direction of
the rotary encoder) by the value specified under
POW:STEP:INCRement. With remote control, the level is set
directly with the command POW:MAN.
STEP
Each trigger triggers one sweep step only. The level increases by
the value entered under :SWEep:POWer:STEP:LOGarithmic.
AUTO
<Mode>
*RST:
SWE:POW:MODE AUTO
selects Mode Auto for a level sweep at RF output.
Example:
[:SOURce<hw>]:SWEep:POWer:POINts <Points>
The command sets the number of steps in a level sweep. The command is linked to the
command:SWEep:POWer:STEP as follows:
POINts = ((STOP - STARt) / STEP:LOG) + 1
If POINTs changes, the value of STEP is adjusted. The STARt and STOP value is retained.
Correlation: The value of :SWE:POW:STEP is adjusted automatically.
Parameters:
float
20dB
<Points>
*RST:
POW:STAR - 30 dBm
sets the start frequency to -30 dBm.
POW:STOP - 10 dBm
sets the stop frequency to -10 dBm.
SWE:POW:POIN 20
sets 20 sweep steps. The sweep step width (STEP) is automati-
cally set to 1 dB.
Example:
[:SOURce<hw>]:SWEep:POWer:SHAPe <Shape>
The command sets the cycle mode for a sweep sequence (shape).
SOURce Subsystem