Remote Control Commands
R&S
®
SMU200A
786Operating Manual 1007.9845.32 ─ 15
Parameters:
string<Ptransition>
STAT:OPER:PTR 32767
all transitions from 0 to 1 in the condition part of the Status:Oper-
ation register cause an entry to be made in the EVENt part.
Example:
SCPI conformUsage:
STATus:OPERation:NTRansition <Ntransition>
The command sets the bits of the NTRansition part of the STATus:OPERation register.
If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in
the EVENt part of the register. The disappearance of an event in the hardware is thus
registered, e.g. the end of an adjustment.
Parameters:
string<Ntransition>
STAT:OPER:NTR 0
a transition from 1 to 0 in the condition part of the Status:Operation
register does not cause an entry to be made in the EVENt part.
Example:
SCPI conformUsage:
STATus:PRESet <Preset>
The command resets the status registers. All PTRansition parts are set to FFFFh (32767),
i.e. all transitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a tran-
sition from 1 to 0 in a CONDition bit is not detected. The ENABle parts of STATus:OPER-
ation and STATus:QUEStionable are set to 0, i.e. all events in these registers are not
passed on.
Parameters:
string<Preset>
STAT:PRES
resets the status registers.
Example:
Event
SCPI conform
Usage:
STATus:QUEStionable:CONDition? <Condition >
The command queries the content of the CONDition part of the STATus:QUEStionable
register. This part contains information on the action currently being performed in the
instrument. The content is not deleted after being read out since it indicates the current
hardware status.
Parameters:
string<Condition >
STATus Subsystem