Remote-Control Commands
R&S
®
SMW-K144
260User Manual 1178.8013.02 ─ 06
[:SOURce<hw>]:BB:NR5G:OUTPut:TDWind:S60K:TRTSamples?
[:SOURce<hw>]:BB:NR5G:OUTPut:TDWind:S240K:TRTSamples?
Queries the number of transition samples.
The next to last block in the command syntax indicates the used SCS and CP combi-
nation.
●
DL: SE<SCS>K, where E indicates the extended CP or for normal CP, the designa-
tion is omitted
Return values:
<TransitionSampl> integer
Range: 0 to 1000
*RST: 0
Example: See Example "Time domain windowing" on page 162.
Usage: Query only
Manual operation: See "Transition BB A Samples" on page 137
[:SOURce<hw>]:BB:NR5G:OUTPut:POWer:MODE <PowerMode>
Sets how the first output is leveled.
Parameters:
<PowerMode> AVG | ACTvsf | PSDConst
AVG
Average RMS power
ACTvsf
Average power in the active subframes
PSDConst
The absolute power of a particular allocation is set by multiplying
the configured power spectral density (PSD) with the bandwidth
of the particular allocation.
*RST: AVG
Example:
SOURce1:BB:NR5G:OUTPut:POWer:MODE PSDConst
SOURce1:BB:NR5G:OUTPut:POWer:RSBW 1000000
Manual operation: See "Power Mode" on page 139
[:SOURce<hw>]:BB:NR5G:OUTPut:POWer:RSBW <RefSystemBW>
Sets the reference bandwidth used by the leveling of the output signal at the first out-
put.
Parameters:
<RefSystemBW> float
Range: 15E3 to 400E6
Increment: 1
*RST: 1E6
Output Commands