Remote Control Commands
R&S
®
SMW200A
1041User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:FREQuency?
Queries the current frequency of the local oscillator at the LO OUT connector.
Return values:
<Frequency> float
Range: 100E3 to 20E9
Increment: 0.01
*RST: 0
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 1038.
Usage: Query only
Manual operation: See "LO OUT Frequency" on page 452
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe <State>
Activates the LO output in path B.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 1038.
Manual operation: See "Out State " on page 452
[:SOURce<hw>]:FREQuency:MODE <Mode>
Sets the frequency mode for generating the RF output signal. The selected mode
determines the parameters to be used for further frequency settings.
SOURce Subsystem