Remote Control Commands
R&S
®
SMW200A
1046User Manual 1175.6632.02 ─ 16
Parameters:
<Start> float
Range: 300kHz to RFmax
Increment: 0.01Hz
*RST: 100 MHz
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See " Start Frequency/Stop Frequency " on page 478
[:SOURce<hw>]:FREQuency:STOP
<Stop>
Sets the stop frequency range for the RF sweep.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Parameters:
<Stop> float
Range: 300kHz to RFmax
Increment: 0.01Hz
*RST: 500 MHz
Default unit: Hz
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See " Start Frequency/Stop Frequency " on page 478
[:SOURce<hw>]:FREQuency:STEP:MODE <Mode>
Defines the type of step size to vary the RF frequency at discrete steps with the com-
mands FREQ UP or FREQ DOWN.
You can define the step mode for each path separately.
Parameters:
<Mode> DECimal | USER
DECimal
Increases or decreases the level in steps of ten.
USER
Increases or decreases the level in increments, set with the
command FREQ:STEP[:INCR].
*RST: DECimal
Example:
// increasing the RF frequency with a step size of 50 KHz
SOURce1:FREQuency:STEP 50E3
SOURce1:FREQuency:STEP:MODE USER
SOURce1:FREQuency:CW UP
Manual operation: See "Variation Active" on page 433