Remote Control Commands
R&S
®
SMW200A
1054User Manual 1175.6632.02 ─ 16
Parameters:
<Type> SINGle | DIFFerential
*RST: SING (if SCONfiguration:EXTernal:PBEHaviour 0) /
- (if SCONfiguration:EXTernal:PBEHaviour 1)
Example:
SOURce1:IQ:OUTPup:ANALog:TYPE DIFFerential
Manual operation: See "I/Q Output Type" on page 158
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:MODE <Mode>
Determines the mode for setting the output parameters.
Parameters:
<Mode> FIXed | VARiable
FIXed
Locks the I/Q output settings
VARiable
Unlocks the settings
*RST: FIXed
Example: See Example "Enabling and configuring the analog output"
on page 1051.
Manual operation: See "Mode" on page 189
[SOURce<hw>]:IQ:OUTPut:LEVel <Level>
Sets the off-load voltage Vp of the analog I/Q signal output.
The value range is adjusted so that the maximum overall output voltage does not
exceed 4V, see "Maximum overall output voltage" on page 186.
Parameters:
<Level> float
Range: 0.04V to 4V (R&S SMW-B10) / 0.04V to 2V
(R&S SMW-B9)
Increment: 1E-4
*RST: 1
Default unit: V
Example: See Example "Enabling and configuring the analog output"
on page 1051.
Manual operation: See "I/Q Level Vp EMF" on page 189
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:COUPling[:STATe] <State>
Couples the bias setting of the I and Q signal components.
SOURce Subsystem