Remote Control Commands
R&S
®
SMW200A
1064User Manual 1175.6632.02 ─ 16
In sweep mode ([:SOURce<hw>]:LFOutput:FREQuency:MODE SWE), the fre-
quency is coupled with the sweep frequency.
Parameters:
<Frequency> float
Range: 0.1 to depends on the installed options
(R&S SMW-K24)
Increment: 0.1
*RST: 1000
Default unit: Hz
Example: See Example "Configuring the LF generator" on page 1062.
Manual operation: See "Frequency" on page 504
[:SOURce<hw>]:LFOutput<ch>:PERiod?
Queries the repetition frequency of the sine signal.
Return values:
<LfSinePeriod> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.001
Default unit: s
Example: See Example "Configuring the LF generator" on page 1062.
Usage: Query only
[:SOURce<hw>]:LFOutput:FREQuency:MANual <Manual>
Sets the frequency of the subsequent sweep step if LFO:SWE:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any value within the setting range, where:
STARt is set with [:SOURce<hw>]:LFOutput:FREQuency:
STARt
STOP is set with [:SOURce<hw>]:LFOutput:FREQuency:
STOP
Range: STARt to STOP
Increment: 0.1
*RST: 1000
Example: See Example "Setup an LF sweep" on page 1061.
Manual operation: See " Current Frequency" on page 478
SOURce Subsystem