Remote Control Commands
R&S
®
SMW200A
1081User Manual 1175.6632.02 ─ 16
See also "Learn List Mode Data list processing mode" on page 473.
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Usage: Event
Manual operation: See " Learn List Mode Data " on page 485
[:SOURce<hw>]:LIST:MODE <Mode>
Sets the list mode.
The instrument processes the list according to the selected mode and trigger source,
see LIST:TRIG:SOUR AUTO, SING or EXT.
Parameters:
<Mode> AUTO | STEP
AUTO
Each trigger event triggers a complete list cycle.
STEP
Each trigger event triggers only one step in the list processing
cycle. The list is processed in ascending order.
In this mode, you can select between LIST:TRIG:SOUR SING or
EXT.
*RST: AUTO
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Manual operation: See "Mode" on page 482
[:SOURce<hw>]:LIST:POWer <Power>
Enters the level values in the selected list. The number of level values must corre-
spond to the number of frequency values. Existing data is overwritten.
Parameters:
<Power> <Power#1>{, <Power#2>, ...} | block data
You can either enter the data as a list of numbers, or as binary
block data.
The list of numbers can be of any length, with the list entries
separated by commas.
In binary block format, 8 (4) bytes are always interpreted as a
floating-point number with double accuracy.
See also :FORMat[:DATA].
Range: depends on the installed options
Default unit: dBm
Example: See Example "Create an RF list and activate the list mode"
on page 1074.
Manual operation: See "Edit List Mode Data" on page 487