Loading...
Remote Control Commands
R&S
®
SMW200A
1087User Manual 1175.6632.02 ─ 16
Manual operation: See "Select Source/Select Destination" on page 491
[:SOURce<hw>]:LIST:DEXChange:AFILe:SEParator:COLumn <Column>
Selects the separator between the frequency and level column of the ASCII table.
Parameters:
<Column> TABulator | SEMicolon | COMMa | SPACe
*RST: COMMa
Example: See Example "List mode data exchange" on page 1077.
Manual operation: See "ASCII File Settings" on page 490
[:SOURce<hw>]:LIST:DEXChange:AFILe:SEParator:DECimal <Decimal>
Sets "." (decimal point) or "," (comma) as the decimal separator used in the ASCII data
with floating-point numerals.
Parameters:
<Decimal> DOT | COMMa
*RST: DOT
Example: See Example "List mode data exchange" on page 1077.
Manual operation: See "ASCII File Settings" on page 490
[:SOURce<hw>]:LIST:DEXChange:MODE <Mode>
Determines the import or export of a list.
Specify the source or destination file with the command [:SOURce<hw>]:LIST:
DEXChange:SELect.
Parameters:
<Mode> IMPort | EXPort
*RST: IMPort
Example: See Example "List mode data exchange" on page 1077.
Manual operation: See " Mode " on page 490
[:SOURce<hw>]:LIST:DEXChange:SELect <Filename>
Selects the ASCII file for import or export, containing a list.
Parameters:
<Filename> string
Filename or complete file path; file extension can be omitted.
Example: See Example "List mode data exchange" on page 1077.
Manual operation: See "Select Source / Select ASCII Destination" on page 491
SOURce Subsystem