Remote Control Commands
R&S
®
SMW200A
1098User Manual 1175.6632.02 ─ 16
Parameters:
<LevMode> NORMal | LOWNoise | LOWDistortion
NORMal
Supplies the RF signal with the standard power level of the
instrument.
LOWNoise
Supplies a very low noise sinewave signal.
LOWDistortion
Supplies a very pure sinewave signal.
*RST: NORMal
Example:
SOURce1:POWer:LMODe LOWD
Sets low distortion mode. The instrument reduces distortions of
the RF signal to a minimum.
Manual operation: See " Mode " on page 436
[:SOURce<hw>]:POWer:MANual
<Manual>
Sets the level for the subsequent sweep step if SWE:POW:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any level within the setting range, where:
STARt is set with [:SOURce<hw>]:POWer:STARt
STOP is set with [:SOURce<hw>]:POWer:STOP
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate]:OFFSet
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01
Default unit: dBm
Example: See Example "Setup an RF frequency or power sweep"
on page 1107
Manual operation: See " Current Level" on page 479
[:SOURce<hw>]:POWer:MODE <Mode>
Selects the operating mode of the instrument to set the output level.
SOURce Subsystem