Remote Control Commands
R&S
®
SMW200A
1103User Manual 1175.6632.02 ─ 16
Parameters:
<Amplitude> float
The following settings influence the value range:
OFFSet set with the command [:SOURce<hw>]:POWer[:
LEVel][:IMMediate]:OFFSet
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command [:SOURce<hw>]:POWer:STEP[:INCRement].
Range: (Level_min + OFFSet) to (Level_max + OFFStet)
*RST: -30
Default unit: dBm
Example:
POWer -30
Sets the RF level
Example: See also [:SOURce<hw>]:POWer:STEP:MODE on page 1100.
Manual operation: See "Amplitude" on page 435
13.18.16 SOURce:ROSCillator Subsystem
The SOURce:ROSCillator subsystem contains the commands for setting the exter-
nal and internal reference frequency.
The commands of this subsystem are not affected by an instrument reset (*RST
on page 772).
Example: Configuring the reference oscillator
// Using 10 MHz external reference source
SOURce:ROSCillator:SOURce EXT
SOURce:ROSCillator:EXTernal:RFOFf:STATe 1
SOURce:ROSCillator:EXTernal:FREQuency 10MHZ
SOURce:ROSCillator:EXTernal:SBANdwidth WIDE
SOURce:ROSCillator:INTernal:ADJust:STATe 0
[:SOURce]:ROSCillator:SOURce................................................................................... 1104
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]...........................................................1104
[:SOURce]:ROSCillator:EXTernal:FREQuency................................................................1104
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable................................................. 1104
[:SOURce]:ROSCillator:EXTernal:SBANdwidth............................................................... 1105
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE.......................................................1105
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue............................................................1105
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]..........................................................1106
SOURce Subsystem