EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 1120

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1105User Manual 1175.6632.02 ─ 16
Example: See Example "Configuring the reference oscillator"
on page 1103.
Manual operation: See "Variable Reference Frequency" on page 447
[:SOURce]:ROSCillator:EXTernal:SBANdwidth <SBandwidth>
Selects the synchronization bandwidth for the external reference signal.
Parameters:
<SBandwidth> WIDE | NARRow
NARRow
The synchronization bandwidth is a few Hz (with option
R&S SMW-B22) or approximately 20 Hz otherwise.
WIDE
The synchronization bandwidth is a 100 Hz (with option
R&S SMW-B22) and approximately 750 Hz otherwise.
*RST: n.a. (factory preset)
Example: See Example "Configuring the reference oscillator"
on page 1103.
Manual operation: See "Synchronization Bandwidth" on page 447
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE <OutpFreqMode>
Selects the mode for the determination and output of the reference frequency.
Parameters:
<OutpFreqMode> DER10M | SAME
*RST: n.a. (factory preset: DER10M)
Example: See Example "Configuring the reference oscillator"
on page 1103.
Manual operation: See "Reference Frequency Output" on page 447
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue <Value>
Specifies the frequency correction value (adjustment value).
Parameters:
<Value> integer
*RST: 0
Example: See [:SOURce]:ROSCillator[:INTernal]:ADJust[:
STATe] on page 1106
Manual operation: See "Adjustment DAC Value" on page 448
SOURce Subsystem

Table of Contents

Related product manuals