Remote Control Commands
R&S
®
SMW200A
1112User Manual 1175.6632.02 ─ 16
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger event triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually by input of the frequencies with the command [:
SOURce<hw>]:FREQuency:MANual.
STEP
Each trigger event triggers one sweep step. The frequency
increases by the value entered with [:SOURce<hw>]:SWEep[:
FREQuency]:STEP[:LINear] (linear spacing) or [:
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
(logarithmic spacing).
*RST:
AUTO
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See "Mode" on page 476
[:SOURce<hw>]:SWEep[:FREQuency]:POINts <Points>
Sets the number of steps within the RF frequency sweep range.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Two separate POINts values are used for linear or logarithmic sweep spacing ([:
SOURce<hw>]:SWEep[:FREQuency]:SPACing LIN | LOG). The command always
affects the currently set sweep spacing.
Parameters:
<Points> integer
Range: 2 to Max
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing <Spacing>
Selects the mode for the calculation of the frequency intervals, with which the current
frequency at each step is increased or decreased.
The keyword [:FREQuency] can be omitted; then the command is SCPI-compliant.
SOURce Subsystem