Remote Control Commands
R&S
®
SMW200A
1114User Manual 1175.6632.02 ─ 16
[:SOURce<hw>]:SWEep:POWer:RETRace <State>
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace <State>
Activates that the signal changes to the start frequency value while it is waiting for the
next trigger event.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See "Retrace" on page 479
[:SOURce<hw>]:SWEep:POWer:RUNNing?
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?
Queries the current sweep state.
Return values:
<State> 0 | 1 | OFF | ON
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Usage: Query only
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic <Logarithmic>
Sets a logarithmically determined step width for the RF frequency sweep. The value is
added at each sweep step to the current frequency.
See Chapter 7.6.2.1, "Correlating Parameters in Sweep Mode", on page 467.
Parameters:
<Logarithmic> float
The unit is mandatory.
Range: 0.01 to 100
Increment: 1E-3
*RST: 1
Default unit: PCT
Example: See Example "Setup an RF frequency or power sweep"
on page 1107.
Manual operation: See " Step Linear/Step Logarithmic " on page 480
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step width for linear sweeps.