Remote Control Commands
R&S
®
SMW200A
1133User Manual 1175.6632.02 ─ 16
Example: See :SYSTem:SPECification? on page 1130.
Usage: Query only
:SYSTem:SPECification:VERSion:FACTory?
Queries the data sheet version of the factory setting.
Return values:
<Version> string
Example: See :SYSTem:SPECification? on page 1130.
Usage: Query only
Manual operation: See "Versions" on page 1171
:SYSTem:SRData?
Queris the SCPI recording data from the internal file.
This feature enables you to transfer an instrument configuration to other test environ-
ments, as e.g. laboratory virtual instruments.
Return values:
<FileData> block data
Example:
SYSTem:SRData?
// #3118:SOURce1:ROSCillator:SOURce EXT
:SOURce1:FREQuency:CW 4000000000
:SOURce1:FREQuency:OFFSet 1000000
:SOURce1:AM1:STATe 1
:OUTPut1:STATe 1
Usage: Query only
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete> 0 | 1 | OFF | ON
*RST: 0
Example:
SYST:STAR:COMP?
Response: 1
the startup of the instrument is completed.
Usage: Query only
:SYSTem:DATE <Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
SYSTem Subsystem