Remote Control Commands
R&S
®
SMW200A
1134User Manual 1175.6632.02 ─ 16
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<Year> integer
<Month> integer
Range: 1 to 12
<Day> integer
Range: 1 to 31
Example:
:SYSTem:DATE?
// 2016,05,01
Manual operation: See "Date" on page 1154
:SYSTem:NTP:HOSTname <NTPName>
Sets the address of the NTP server. You can enter the IP address, or the hostname.
Parameters:
<NTPName> string
Example:
SYSTem:NTP:HOSTname "pool.ntp.org"
Manual operation: See "NTP Address" on page 1155
:SYSTem:NTP:STATe <UseNtpState>
Activates clock synchronization via NTP.
Parameters:
<UseNtpState> 0 | 1 | OFF | ON
*RST: 0
Example:
SYSTem:NTP:STATe 1
Manual operation: See "Use Time from NTP Server" on page 1155
:SYSTem:TIME <Hour>, <Minute>, <Second>
Queries or sets the time for the instrument-internal clock.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<Hour> integer
Range: 0 to 23
<Minute> integer
Range: 0 to 59
SYSTem Subsystem