Remote Control Commands
R&S
®
SMW200A
1135User Manual 1175.6632.02 ─ 16
<Second> integer
Range: 0 to 59
Example:
SYSTem:TIME?
// 10,27,14
Manual operation: See "Time" on page 1155
:SYSTem:TIME:ZONE <TimeZone>
Sets the timezone. You can query the list of the available timezones with :SYSTem:
TIME:ZONE:CATalog?.
Parameters:
<TimeZone> string
Manual operation: See "Timezone" on page 1155
:SYSTem:TIME:ZONE:CATalog?
Querys the list of available timezones.
Return values:
<Catalog>
Usage: Query only
Manual operation: See "Timezone" on page 1155
:SYSTem:UPTime?
Queries the up time of the operating system.
Return values:
<UpTime> "<ddd.hh:mm:ss>"
Example:
SYSTem:UPTime?
Response: "0.08:11:00"
Usage: Query only
:SYSTem:BIOS:VERSion?
Queries the BIOS version of the instrument.
Return values:
<Version> string
Example:
SYST:BIOS:VERS?
queries the BIOS version.
Response: 123456
Usage: Query only