EasyManua.ls Logo

Rohde & Schwarz SMW200A - External Instrument; I;Q Connection; Remote Connection Status

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Routing and System Configuration
R&S
®
SMW200A
152User Manual 1175.6632.02 ─ 16
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:DIRection? on page 845
:SCONfiguration:EXTernal:FADer<ch>:DIRection? on page 845
:SCONfiguration:EXTernal:BBMM<ch>:DIRection? on page 845
:SCONfiguration:EXTernal:IQOutput<ch>:DIRection? on page 845
External Instrument
Accesses a dialog with settings that help you to establish the connection to the exter-
nal instruments, see Chapter 3.4.3.2, "External Instrument Configuration Settings",
on page 154.
Note:
Each external instrument must be connected via a LAN or USB to the R&S SMW.
In most test setups, all instruments must use a common reference frequency signal,
either from an external common reference source or using the distributed reference
frequency of the R&S SMW.
See also:
Chapter 3.9.5, "How to Cable the Instruments in MIMO Test Setups", on page 197
Chapter 3.9.6, "How to Connect and Configure External Instruments",
on page 198.
I/Q Connection
Depends on the interface type.
Analog I/Q interfaces:
Selects the type of the I/Q output signal
Digital I/Q interfaces:
Indicates the status of the I/Q connection of the digital interfaces.
Indication
Designation
A green LED and
an icon
The DIG I/Q interface is used as an input connector
A yellow LED and
an icon
The DIG I/Q interface is used as an output connector
An icon without an
LED
An external instrument is connected to the DIG I/Q interface, but
no direction is specified
No icon displayed
No connection
"Single Ended"/ "Differential" Selects the type of the I/Q output signal at the I/Q output connec-
tor
Works like the parameter "I/Q Output Type" on page 158.
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:IQConnection:STATe? on page 846
:SCONfiguration:EXTernal:FADer<ch>:IQConnection:STATe? on page 846
:SCONfiguration:EXTernal:BBMM<ch>:IQConnection:STATe? on page 846
Remote Connection Status
Indicates the status of the remote connection. Click the status indicator to toggle the
remote connection status.
System Configuration Settings

Table of Contents

Related product manuals