EasyManua.ls Logo

Rohde & Schwarz SMW200A - Page 173

Rohde & Schwarz SMW200A
1312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Routing and System Configuration
R&S
®
SMW200A
158User Manual 1175.6632.02 ─ 16
Remote command:
:SCONfiguration:EXTernal:REMote:ADD on page 850
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO? on page 852
:SCONfiguration:EXTernal:FADer<ch>:REMote:INFO? on page 852
:SCONfiguration:EXTernal:BBMM<ch>:REMote:INFO? on page 852
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO? on page 852
I/Q Output Type
Selects the type of output signal.
If R&S SMW-B9 is installed, the differential signal output can be activated in "I/Q Ana-
log A" block only. Single-ended and differential signals cannot be output at the same
time.
The provided parameters in the "I/Q Analog Outputs" dialog depend on the selected
output mode.
"Differential"
If "RF Envelope > On"
The inverted envelope signal Ē is output at the I BAR connectors.
If "RF Envelope > Off"
The analog I/Q signal components are output at the I/Q BAR con-
nectors.
"Single-Ended"
If "RF Envelope > On"
The envelope signal E is output at the I connectors.
If "RF Envelope > Off"
Single-ended output at the I/Q connectors.
You can define a bias between the output signal and ground.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE on page 1053
Apply
Confirms the settings and displays the external instrument in the "External RF and I/Q"
dialog.
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:REMote:ISELect on page 851
:SCONfiguration:EXTernal:FADer<ch>:REMote:ISELect on page 851
:SCONfiguration:EXTernal:BBMM<ch>:REMote:ISELect on page 851
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:ISELect on page 851
Apply and Connect
Confirms the settings and triggers the connection establishment. In the connected
instruments, the status of this active remote connection is displayed by the "Remote"
indication.
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:REMote:ISELect on page 851
:SCONfiguration:EXTernal:CODer<ch>:REMote:CONNect on page 851
:SCONfiguration:EXTernal:FADer<ch>:REMote:ISELect on page 851
:SCONfiguration:EXTernal:FADer<ch>:REMote:CONNect on page 851
:SCONfiguration:EXTernal:BBMM<ch>:REMote:ISELect on page 851
System Configuration Settings

Table of Contents