Signal Routing and System Configuration
R&S
®
SMW200A
177User Manual 1175.6632.02 ─ 16
Settings:
Signal Monitoring........................................................................................................ 177
Signal Monitoring
The R&S SMW monitors constantly the input signal (Crest Factor and Peak Level) and
indicates an overflow status, if the entered "Peak Level" (in dB full scale) value does
not correspond with the real value.
Note: The overflow status is indicated as long as the "Peak Level" value is not correc-
ted.
To avoid overflow effects, enable automatic measurement with the "Auto Level Set"
function.
"Overflow/No Overflow"
Indicates whether the input is overloaded or not. An additional indica-
tion is displayed in the block diagram close to the corresponding con-
nector.
Remote command:
[:SOURce<hw>]:BBIN:OLOad:STATe? on page 869
"Overflow Hold"
Indicates an overload since the last reset for evaluating the measure-
ment.
To reset the state, use one of the following:
Select "Reset"
Start a new measurement with the "Auto Level Set" function
Change the value of one of the monitored parameters.
Remote command:
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe? on page 869
"Reset"
Resets the "Overload Hold" indication.
Remote command:
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe? on page 869