Configuring the Internal Baseband Source
R&S
®
SMW200A
246User Manual 1175.6632.02 ─ 16
Example:
This example shows schematically the distribution of the trigger signal in instrument
with the following configuration:
"System Configuration > Entity = 1", 4x4 MIMO configuration (1x4x4) and "Base-
band Source Config > Separate".
"Baseband A > ... > Trigger In > Trigger Source > Ext. Global Trigger 1"
Each of the "Baseband B/C/D" adopts the same trigger source automatically.
Figure 4-11: Distribution of the external trigger signal in the coupled trigger settings mode
"Baseband A > ... > Trigger In > Trigger Mode > Retrigger"
The "Baseband B/C/D" adopt the same trigger mode automatically.
Figure 4-12: Simultaneous triggering of two basebands in coupled trigger settings mode
** = this example assumes identical basebands and shows only the first two of them
Common Functions and Settings in the Baseband Domain