Configuring the Internal Baseband Source
R&S
®
SMW200A
257User Manual 1175.6632.02 ─ 16
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut
on page 906
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:SYNChronize:OUTPut
on page 956
[:SOURce<hw>]:BB:MCCW:TRIGger:EXTernal:SYNChronize:OUTPut
on page 996
External / Trigger Inhibit ← Trigger Settings Common to All Basebands
For external trigger signal or trigger signal from the other path, sets the duration with
that any following trigger event is suppressed. In "Retrigger" mode for example, a new
trigger event does not cause a restart of the signal generation until the specified inhibit
duration does not expire.
For more information, see "Impact of the Additional Trigger Settings" on page 241.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal]:INHibit on page 910
[:SOURce<hw>]:BB:DM:TRIGger:OBASeband:INHibit on page 908
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:INHibit on page 958
[:SOURce<hw>]:BB:ARBitrary:TRIGger:OBASeband:INHibit on page 957
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal]:INHibit on page 998
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:INHibit on page 996
(External) Delay Unit
Determine whether the trigger delay is expressed in samples or directly defined as a
time period (seconds).
To specify the delay, use the (Specified) External Delay/(Specified) Trigger Delay
The parameter Actual Trigger Delay/Actual External Delay displays the delay conver-
ted in time.
See also "Specifying delay and inhibit values in time units" on page 243.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:DELay:UNIT on page 908
[:SOURce<hw>]:BB:ARBitrary:TRIGger:DELay:UNIT on page 957
(Specified) External Delay/(Specified) Trigger Delay
The name of the parameter and the units the delay is expressed in, changes depend-
ing on the parameter (External) Delay Unit.
Delays the trigger event provided by:
●
The external trigger source
●
The other path
●
The other basebands (internal trigger), if common trigger settings are used.
●
Other instrument in master-slave mode
Use the settings to:
●
Synchronize the instrument with the device under test (DUT) or other external devi-
ces
●
Postpone the signal generation start in the basebands compared to each other
●
Compensate for signal delay and align the signals in time
Common Functions and Settings in the Baseband Domain