Configuring the Signal in the RF Domain
R&S
®
SMW200A
432User Manual 1175.6632.02 ─ 16
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
The display has been disabled for security reasons.
See:
Annotation Frequency
Annotation Amplitude
The display is disabled when:
List mode is running, see Chapter 7.6, "Varying the RF Signal in List or Sweep
Mode", on page 457.
Digital only signals are generated, see the user manual "R&S SMW-K551 Gen-
eration of Digital "Slow IQ" Signals"
Remote command:
[:SOURce<hw>]:FREQuency[:CW|FIXed] on page 1042
Offset
Sets the frequency offset.
This value represents the frequency shift of a downstream instrument, like for example
a mixer.
The "Frequency" value displayed in the status bar is the resulting frequency, as it is at
the output of the downstream instrument. The frequency at the R&S SMW RF output is
not changed.
See Chapter 7.3.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 440.
Remote command:
[:SOURce<hw>]:FREQuency:OFFSet on page 1044
Multiplier
Sets the multiplication factor for the RF frequency.
This value represents the multiplication factor of a downstream instrument, like for
example a multiplier.
The "Frequency" value displayed in the status bar is the resulting frequency, as it is at
the output of the downstream instrument. The frequency at the R&S SMW RF output is
not changed.
See Chapter 7.3.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 440.
Remote command:
[:SOURce<hw>]:FREQuency:MULTiplier on page 1044
Converter Mode
Selects the mode for calculating the output frequency of a downstream converter.
Configuring RF Frequency and Level