Configuring the Signal in the RF Domain
R&S
®
SMW200A
445User Manual 1175.6632.02 ─ 16
"System Config" > "Setup" > "Reference Freq/LO Coupling ..."
In the "Reference Freq/LO Coupling" tab, you can select the signal source of the
reference frequency, enter and adjust the frequency of an external reference sig-
nal, and configure the Local Oscillator Coupling.
Instruments equipped with the latest SSYN assembly (part no. 1412.6667.02) allows
you to select either 10 MHz or a variable external reference frequency, whereas the
instruments with the previous SSYN assembly (part no. 1141.4220.02) support only 5,
10 or 13 MHz, see Table 7-2.
For an overview of the installed hardware options, see Chapter 14.3.8.1, "Hardware
Configuration Settings", on page 1168
Reference Oscillator