Configuring the Signal in the RF Domain
R&S
®
SMW200A
485User Manual 1175.6632.02 ─ 16
List Range from/to
Defines an index range in the current list by setting the start and stop index. Only the
values in the selected index range are processed in list mode, all other list entries are
ignored.
Remote command:
[:SOURce<hw>]:LIST:INDex:STARt on page 1080
[:SOURce<hw>]:LIST:INDex:STOP on page 1080
Run Mode
Selects whether the instrument generates the signal by processing the list directly, or
by learning and replaying the data.
"Learned (Frozen RF Settings)"
Generates the signal with the previously learned and stored data from
the temporary memory.
See "Learn List Mode Data list processing mode" on page 473.
"Live"
Generates the signal directly from the database.
The instrument reads the pairs of values from the list, calculates the
hardware settings and generates the signal immediately.
See "Live list processing mode" on page 474.
Remote command:
[:SOURce<hw>]:LIST:RMODe on page 1082
Learn List Mode Data
Reads the settings from the list data table and calculates the associated hardware set-
tings. The learned data is stored in the temporary memory.
See also "When to use the "Learn List Mode Data" function" on page 473.
Remote command:
[:SOURce<hw>]:LIST:LEARn on page 1080
7.6.6.3 Import/Export Settings
Access:
1. Select "RF" > "Sweep/List" > "List mode".
Varying the RF Signal in List or Sweep Mode