File and Data Management
R&S
®
SMW200A
626User Manual 1175.6632.02 ─ 16
Example:
Figure 10-2: Labeled parameters show that the value deviates from its default
Frequency = changed (default = 1 GHz)
Offset = changed (default = 0 kHz)
Multiplier = unchanged
Variation active = changed (default = enabled)
Variation step = unchanged
10.2.3 How to Recall User Settings Automatically after Preset
You can define the settings that are restored when you preset the instrument.
1. Configure the settings as required. Save them as described in Chapter 10.4.2,
"How to Save and Recall Instrument Settings", on page 632.
2. Save the settings as a file with the predefined filename UserPreset.savrcltxt.
Save this file in the directory /var/user/.
The filename UserPreset.savrcltxt and the directory /var/user/ are man-
datory.
Now when you press the PRESET key or send the *RST command to the instru-
ment, the defined settings are restored.
An "Info" message confirms, that a file with user-defined preset setting is loaded.
Restoring the (Default) Instrument Configuration