Getting Started
R&S
®
SMW200A
59User Manual 1175.6632.02 ─ 16
●
Connection of peripherals such as mouse or keyboard
●
Connection of memory stick for file transfer
●
Firmware update
Note: Further USB interface type A (host USB) and a USB interface type B (USB
DEVICE) are available on the rear panel.
See also Chapter 2.1.3, "Connecting USB Devices", on page 44.
SENSOR
Connector for R&S NRP sensors.
A power sensor is connected to the R&S SMW by inserting the male connector. To dis-
connect, pull the connector by its sleeve. You cannot disconnect the sensor simply by
pulling at the cable or the rear part of the connector.
The R&S SMW supports the use of R&S NRP power sensors in various ways including
the use as a power viewer.
See Chapter 7.8.4, "Using Power Sensors", on page 540 and Chapter 7.8.3, "User
Correction", on page 529.
USER X
Multipurpose connectors. The input/output signal has to be defined.
The Table 2-6 lists the signals assigned to the USER X connectors in the default
instrument state.
Table 2-6: Default configuration of the USER X connectors
USER connector Direction Default assigned signal
1 Output Baseband A Marker 1
2 Output Baseband A Marker 2
3 Input Global Trigger 1
A dedicated LED indicates the connector status:
●
Green: an input connector
●
Yellow: an output connector
●
No light: the connector is not active
See also Chapter 11.2, "Configuring the Local and Global Connectors", on page 661
and Chapter 11.2.4, "RF Connectors Settings", on page 667.
RF A/RF B
Outputs for RF signal of path A and path B.
NOTICE! Maximum input levels. Do not overload the RF output. The maximum permis-
sible back-feed is specified in the data sheet.
The connector type depends on the installed frequency option, see Table 2-7.
Instrument Tour