Network Operation and Remote Control
R&S
®
SMW200A
738User Manual 1175.6632.02 ─ 16
The instrument response is 4000000000. The value corresponds to the frequency
in Hz.
While remote control is active, the "Remote" icon in the status bar indicates that the
instrument is in remote control mode. Currently ongoing communication (data
transfer) is indicated by green colored arrows in the icon.
Additionally, the "Connections" softkey in the taskbar shows the currently estab-
lished remote connections. The operation via the front and touch panel or via
mouse and keyboard are locked, allowing a remote control program to be per-
formed without interruption.
On the display, keys and entry fields are grayed out and cannot be activated or
modified, but you can still open dialogs, for example to verify settings.
5. To disable the access to the dialogs, use the command SYST:KLOC ON.
6. To prevent unintentional return to manual operation, use the command &LLO.
See also Chapter B.1.2, "LAN Interface Messages", on page 1192.
The instrument switches to "Remote LLO" state. The SETUP key is disabled.
7. To enable the SETUP key, use the command >R.
8. To return to manual operation, see Chapter 12.10, "How to Return to Manual Oper-
ation", on page 742.
Tip: Switching from manual operation to remote control and vice versa does not
affect the other instrument settings.
How to Set Up a Remote Control Connection