Remote Control Commands
R&S
®
SMW200A
775User Manual 1175.6632.02 ─ 16
Example:
SOUR:PRES
Presets all settings that are related to signal path
Usage: Event
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
●
The PRESET key
●
The *RST command
For an overview of the settings affected by the preset function, see Table 10-1
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage: Setting only
:SYSTem:FPReset
Triggers an instrument reset to the original state of delivery.
Example:
SYST:FPR
All instrument settings (also the settings that are not currently
active) are reset to the factory values.
Usage: Event
Manual operation: See "Execute Factory Preset" on page 625
13.7 MMEMory Subsystem
The MMEMory subsystem (Mass MEMory) contains the commands for managing files
and directories as well as for loading and storing complete instrument settings in files.
Mass storage location
Without any additional measures, the R&S SMW stores user files on the internal mem-
ory, or if selected, on a memory stick.
Both, the user directory /var/user on the internal memory or the /var/usb direc-
tory, can be used to preserve user-defined data. Any directory structure can be cre-
ated.
The /var/volatile directory serves as a RAM drive and can be used to protect
sensitive information. The data is available temporarily.
Default storage location
The R&S SMW stores user data in the user directory.
MMEMory Subsystem