Remote Control Commands
R&S
®
SMW200A
789User Manual 1175.6632.02 ─ 16
:CALibration:ALL:WARN.................................................................................................790
:CALibration<hw>:LEVel:BWIDth.....................................................................................790
:CALibration<hw>:ALL:DATE?........................................................................................ 790
:CALibration<hw>:ALL:TEMP?........................................................................................ 791
:CALibration<hw>:DEBug............................................................................................... 791
:CALibration<hw>:FMOFfset[:MEASure]?.........................................................................791
:CALibration<hw>:FREQuency[:MEASure]?......................................................................791
:CALibration<hw>:IQModulator:LOCal?............................................................................792
:CALibration<hw>:LEVel:HACCuracy[:STATe].................................................................. 792
:CALibration<hw>:FREQuency:CONVerter:EXTernal?.......................................................792
:CALibration<hw>:LEVel:LOCal?..................................................................................... 793
:CALibration<hw>:LEVel:STATe......................................................................................793
:CALibration<hw>:LEVel[:MEASure]?...............................................................................793
:CALibration<hw>:LOSCillator:COUPling:LOCal?..............................................................793
:CALibration:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not need external measuring equipment.
Note: If an external frequency converter R&S SZU is connected, the internal adjust-
ments are blocked, and the R&S SMW generates an error message.
Query parameters:
<Force> string
Return values:
<Measure> 0 | 1 | OFF | ON
Example:
CAL:ALL:MEAS?
//"0"
// Executes the adjustments of all instrument functions.
// When completed, it indicates that the adjustment
// has been performed successfully.
Usage: Query only
Manual operation: See "Adjust All" on page 1161
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date> string
Example:
CAL:DATA:FACT:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Factory Calibration" on page 1169
CALibration Subsystem