Remote Control Commands
R&S
®
SMW200A
790User Manual 1175.6632.02 ─ 16
:CALibration:ROSCillator[:DATA] <Data>
Sets a user-defined calibration value for the internal reference frequency.
Parameters:
<Data> integer
Range: 0 to INT_MAX
*RST: 0
Example: See [:SOURce]:ROSCillator[:INTernal]:ADJust[:
STATe] on page 1106.
:CALibration<hw>:CONTinueonerror
<State>
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
CAL:CONT ON
// Continues calibration after an error
Manual operation: See "Continue Adjustment on Error" on page 1162
:CALibration:ALL:WARN <UncalWarning>
Generates a warning message, if the instrument is in uncalibrated state.
Parameters:
<UncalWarning> 0 | 1 | OFF | ON
*RST: 0
Example:
CALibration:ALL:WARN 1
:CALibration<hw>:LEVel:BWIDth <Bandwidth>
Parameters:
<Bandwidth> AUTO | HIGH
*RST: AUTO
:CALibration<hw>:ALL:DATE?
Queries the date of the most recently executed full adjustment.
Return values:
<Date> string
Example:
CAL:ALL:DATE?
// "2016-01-01"