Remote Control Commands
R&S
®
SMW200A
791User Manual 1175.6632.02 ─ 16
Usage: Query only
Manual operation: See "Last Full Adjustment" on page 1161
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Return values:
<Temperature> string
Example:
CALibration:ALL:TEMP?
// "+39.00 K"
Usage: Query only
Manual operation: See "Temperature Offset To Last Full Adjustment"
on page 1161
:CALibration<hw>:DEBug <State>
Activates logging of the internal adjustments.
Setting parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
CALibration:DEBug 1
Usage: Setting only
Manual operation: See "Log Debug Info" on page 1162
:CALibration<hw>:FMOFfset[:MEASure]?
Starts adjustment of the FM/PhiM modulator.
Return values:
<Measure> 0 | 1 | OFF | ON
Example:
CALibration:FMOFfset:MEASure?
// starts adjustment
// 0
// Adjustment successful
Usage: Query only
:CALibration<hw>:FREQuency[:MEASure]?
Performs all adjustments which affect the frequency.
Return values:
<Measure> 0 | 1 | OFF | ON
CALibration Subsystem